Skip to content

Inaccurate analog input values #31

Description

@Cody-G

This is a bit backwards as I've already submitted the pull request #30 for this, but just for tracking the issue:

We were reading raw 16-bit integer values from the DAQ card, which is not advised with current NI DAQ cards because this value has not been scaled to reflect calibration. For the most part this only introduced a slope error in the voltage - to - digital value conversion, so we didn't notice.

Activity

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment

Metadata

Metadata

Assignees

No one assigned

    Labels

    Type

    No type

    Projects

    No projects

      Milestone

      No milestone

      Relationships

      None yet

      Development

      No branches or pull requests

      Issue actions